C-600
Typical Installation |
|
Mounting Location
-
Choose an
area located near the chemical supply tank, chemical injection point and
electrical supply. Install the pump where it can be easily serviced.
-
Mount the
pump to a secure surface or wall. Wall mount to a solid surface only.
Mounting to drywall with anchors is not recommended.
-
Do not
mount the pump directly over your chemical container. Chemical fumes may
damage the unit. Mount the pump off to the side or at a lower level than
the chemical container opening.
-
Mounting
the pump lower than the chemical container will gravity feed the
chemical into the pump. This "flooded suction" installation can reduce
the time required to prime the pump. Install a shut-off valve, pinch
clamp or other means to halt the gravity feed to the pump during
servicing.
-
Your
solution tank should be sturdy. Keep the tank covered to reduce fumes.
-
Be sure
your installation does not constitute a cross connection with the
drinking water supply. Check your local plumbing codes.
|
|
Note:
All diagrams are strictly for guideline purposes only. Always consult an
expert before installing the Chem-Feed pump on specialized systems. The
Chem-Feed should be serviced by qualified persons only. |
|