With the pump installed under
normal operating conditions, place the suction tubing in a
graduated container.
-
Fill the container with the same
chemical normally injected, and run the pump until all air
is removed. (Prime the pump).
-
Refill the container if
necessary, and with the foot valve in the solution, note the
amount of chemical in the container.
-
Using a stopwatch or timer, run
the pump for a measured amount of time (example: 1 minute).
Note the amount of chemical pumped during that
time. Tip: The longer the testing time, the
more accurate the measurement.
Example:
During your 1 minute calibration
period, say your injector pumped 2.5 ounces.
2.5 ounces per minute
(OZM)

|
 |
Your pump is calibrated to pump
1.17 GPH - or
- 28.08
GPD | |
|
Caution:
Always wear protective
clothing, face shield, safety glasses and gloves when working
on or near your metering pump. Additional precautions
should be taken depending on the solution being pumped. Refer
to MSDS precautions from your solution
supplier. |
|
Note: All diagrams are strictly for
guideline purposes only. Always consult an expert before
installing the metering pump on
specialized systems. The metering pump
should be serviced by qualified persons
only. | |
|